TOS5300_INTERFACE
21
3
Co
m
m
a
n
d
R
e
fe
re
n
ce
AC Withstanding Voltage Test (ACW) Conditions
These are commands for setting the AC withstanding voltage test conditions.
Measurement methods setting (ACW)
SENS:MODE
Sets the measurement method.
Command
SENSe[:ACW]:MODE {RMS|AVE}
SENSe[:ACW]:MODE?
Parameter
Value:
RMS
True rms response (default)
AVE
Mean-value response
Response
Returns the measurement method in <character> format.
Test voltage setting (ACW)
SOUR:VOLT
Sets the test voltage.
Command
SOURce[:ACW]:VOLTage[:LEVel] {<numeric>|MIN|MAX}
SOURce[:ACW]:VOLTage[:LEVel]? {MIN|MAX}
Parameter
Value:
0 to 5.5 k (The default value is 0.)
Unit:
V
Response
Returns the test voltage setting in <NR3> format.
Limit voltage setting (ACW)
SOUR:VOLT:PROT
Sets the limit voltage.
Command
SOURce[:ACW]:VOLTage:PROTection[:LEVel][:UPPer] {<numeric>|MIN|MAX}
SOURce[:ACW]:VOLTage:PROTection[:LEVel][:UPPer]? {MIN|MAX}
Parameter
Value:
0 to 5.5 k (The default value is 5.5 k.)
Unit:
V
Response
Returns the limit voltage in <NR3> format.
* RST
* RCL
* RST
* RCL
* RST
Содержание TOS5300
Страница 8: ...8 TOS5300_INTERFACE This page is intentionally blank ...
Страница 14: ...14 TOS5300_INTERFACE This page is intentionally blank ...
Страница 15: ...Command Reference This chapter explains topics such as command details and registers Remote Control ...
Страница 76: ...76 TOS5300_INTERFACE ...