38
TOS5300_INTERFACE
Querying Measured Values (SEQuence1:ACQuire)
MEAS:VOLT / READ:VOLT
Starts a new measurement and queries the voltage.
Command
MEASure[:ARRay]:VOLTage?
READ[:ARRay]:VOLTage?
Response
Returns the measured voltage in <NR3> format or the measured voltages as a comma-separated
list in <NR3>,<NR3>,... format.
Unit:
V
MEAS:RES / READ:RES
Starts a new measurement and queries the resistance.
Command
MEASure[:ARRay]:RESistance?
READ[:ARRay]:RESistance?
Response
Returns the measured resistance in <NR3> format or the measured resistances as a comma-
separated list in <NR3>,<NR3>,... format.
Unit:
OHM
MEAS:TIME / READ:TIME
Starts a new measurement and queries the elapsed test time.
Command
MEASure[:ARRay]:TIME?
READ[:ARRay]:TIME?
Response
Returns the measured elapsed test time in <NR3> format or the measured elapsed as a comma-
separated list in <NR3>,<NR3>,... format.
Unit:
s
TRIG:SOUR
Sets the trigger source to apply to the SEQuence1 (ACQuire) group.
Command
TRIGger[:SEQuence[1]]:SOURce {IMMediate|BUS|TIMer|TEST}
TRIGger[:SEQuence[1]]:SOURce?
TRIGger[:ACQuire]:SOURce {IMMediate|BUS|TIMer|TEST}
TRIGger[:ACQuire]:SOURce?
Parameter
Value:
IMMediate
Begin measuring immediately. (default).
BUS
Wait for a software trigger (a *TRG, TRIG, or IEEE 488.1 get—Group
Execute Trigger—command), and then begin measuring.
TIMer
Begin measuring when the time set with TRIG:TIM elapses.
TEST
Begin measuring when a test begins.
Response
Returns the trigger source in <character> format.
* RST