32
TOS5300_INTERFACE
Insulation Resistance Test (IR) Conditions
SOUR:IR:VOLT:TIM:STAT
Sets whether to stop testing after the set test time elapses (TIMER ON/OFF).
Command
SOURce:IR:VOLTage:TIMer:STATe {ON|OFF|1|0}}
SOURce:IR:VOLTage:TIMer:STATe?
Parameter
Value:
ON (1)
Testing is stopped after the test time elapses (default).
OFF (0)
Testing is not stopped after the test time elapses.
Response
Returns whether testing is stopped after the test time elapses in <NR1> format.
Judgment wait time settings (IR)
SENS:IR:JUDG:DEL
Sets the time to wait after testing starts before starting judgment (WAIT).
Command
SENSe:IR:JUDGment:DELay {<numeric>|MIN|MAX}
SENSe:IR:JUDGment:DELay? {MIN|MAX}
Parameter
Value:
0.1 to 10.0 (The default value is 0.1.)
Unit:
S
Response
Returns the judgment wait time in <NR3> format.
* RCL
* RST
* RCL
* RST