80
PFX2512_CE
Preparation of the DUT (battery) Connection
Connecting the thermocouple for connector plug
Connect type K thermocouples directly to the connector plug.
Making the cell voltage sensing cable
Connect shielded cables directly to the connector plug.
Thermocouple
Thermocouple 4
Thermocouple 3
Thermocouple 2
Thermocouple 1
The attached
connector plug
1
2
NC
NC
3
4
Insert the wire with 7 mm of
the covering removed here.
Temperature Sensing
connector on the
rear panel
Length up to 5 m
TERM
+
-
+
-
+
-
+
-
1234
The attached
connector plug
Shielding potential
1
2
FG
FG
3
4
Cell voltage sensing cable
(2-conductor shielded wire)
+
-
+
-
+
-
+
-
1
FG
234
FG
TERM
MAX20V
Length up to 5 m
Insert the wire with 7 mm of
the covering removed here.
Voltage sensing connecting terminal 4
Voltage sensing connecting terminal 3
Voltage sensing connecting terminal 2
Voltage sensing connecting terminal 1
Voltage Sensing
connector on the
rear panel