24
PFX2512_CE
Connection and Setting of the Each Equipment
The following example is consisting of the Charge/ Discharge System Controller (PFX2512), DC
power supply (PWR800L), the Electronic load (PLZ1004W), and the DUT (battery).
The direction of leading cables of input / output terminal board can be either right or left angle.
However, wire the all the cables to become same directions as possible. Use the cable whichever it
comes as standard accessory.
Please note the following for the stability of operation, the reduction of noise effect, and the
prevention of the malfunction.
•
Twist the cables between the equipments.
•
Do not entwine or cross of the connected cables between the equipments.
•
Do not entwine the cable with the flat cable which flows, and do not pass through the
clearance between the +DUT cable and the -DUT cable.
The following describes connecting procedure for the entire system. Please refer to the connection
between each equipment.
■
Connection for the chassis terminal
The chassis terminal is used for an earthing terminal as multipurpose usage. It can be applied to the
reduction of noise under the large current operation, and prevention of the malfunction. Even if the
cable come in contact with the chassis by any chance, it can prevent the electric shock accident. To
secure the safety operation of the charge and discharge testing, it is recommended to earthing the
chassis terminal.
For the ground cable of the chassis terminal, use the cable with higher capacity of the maximum
current necessary for the charge and discharge testing.
WARNING
Possible electric shock. It is not concerned of any sequence of the connecting equipments,
however, the DUT (battery) must be connected at the final connection and it must be
connected from the side of PFX2512.
–
+
PLZ1004W
PWR800L
PFX2512
– DC
– DC INPUT +
J1
J1
To PWR +
To PWR
–
To DUT +
To DUT
–
To PLZ +
To PLZ
–
To PWR J1
To PLZ J1
26-conductor flat cable
(Accessory)
Cable with crimp
terminal (Accessory)
Chassis terminal
Thermistor
(Accessory)
20-conductor flat cable
(Accessory)
–
+
DUT cable
DUT cable
Voltage
sensing cable
Temperature
sensing cable
DUT (battery)
+
–
The dedicated load cable (TL08-PFX) is available
as an optional item.
*
*
*
Example of connecting figure of the charge / discharge system with the DUT (battery)
Содержание PFX2515
Страница 10: ...10 PFX2512_CE This page is intentionally blank ...
Страница 11: ...General Description This chapter describes the outline of product the connectable equipments and the options ...
Страница 58: ...58 PFX2512_CE This page is intentionally blank ...
Страница 59: ...Specification This chapter contains the PFX2512 specifications and outline drawings ...
Страница 86: ...86 PFX2512_CE This page is intentionally blank ...
Страница 100: ...100 PFX2512_CE This page is intentionally blank ...
Страница 103: ......