Chapter 5: Troubleshooting
To troubleshoot the oscilloscope
75
Primary Trouble Isolation Flowchart
Primary Trouble Isolation
Perform power-up.
Check display.
On
screen display
problems
?
Run scope self tests.
Does
self test
pass?
Check front panel
response.
Do
knobs and keys test
OK?
System works; do
performance tests.
End
Go to 'No Display
Debug.'
Go to 'Keyboard
Troubleshooting.'
Does
LED test
OK?
Check Self Calibration.
Does
self calibration test
pass?
Go to 'Acquisition
Troubleshooting.'
No
No
No
Yes
A
B
E
F
G
H
54830F01a
Check processor
temperature.
C
Is
temperature
OK?
Yes
Check for fan fail
message.
Does
fan fail?
No
Replace fan.
Yes
D
Go to 'Acquisition
Troubleshooting.'
Yes
No
Yes
No
Yes
Yes
Replace
motherboard.
Replace keypad
(includes LEDs).
No
Содержание Infiniium 8000 Series
Страница 6: ...6...
Страница 31: ...Chapter 2 Preparing for Use Digital probe lead set 31 Data Sheets Demonstrations Catalogs...
Страница 66: ...Chapter 3 Testing Performance Performance Test Record 66...
Страница 140: ...Chapter 7 Replaceable Parts Exploded Views 140 Exploded Views Front Frame and Front Panel...
Страница 141: ...Chapter 7 Replaceable Parts Exploded Views 141 Fan and Acquisition Assembly...
Страница 142: ...Chapter 7 Replaceable Parts Exploded Views 142 Power Supply and PC Motherboard Without Option 017...
Страница 143: ...Chapter 7 Replaceable Parts Exploded Views 143 Power Supply and PC Motherboard With Option 017...
Страница 144: ...Chapter 7 Replaceable Parts Exploded Views 144 Sleeve and Accessory Pouch...
Страница 150: ...Chapter 8 Theory of Operation 150 Figure 8 1 Oscilloscope Block Diagram...
Страница 158: ...158...
Страница 160: ...160 Index...