Chapter 2: Preparing for Use
To verify basic oscilloscope operation
34
To verify basic oscilloscope operation
1
Connect an oscilloscope probe to channel 1.
2
Attach the probe to the probe compensation output on the front panel of the
oscilloscope.
Use a probe grabber tip so you do not need to hold the probe. The probe compensation
output is marked with a square wave symbol.
Figure 2-16
3
Press the Default Setup key on the front panel.
The display will pause momentarily while the oscilloscope is configured to its default
settings.
4
Press the Autoscale key on the front panel.
The display will pause momentarily while the oscilloscope adjusts the sweep speed and
vertical scale. You should then see a square wave with an amplitude of approximately
1.1 Vpp at about 700 to 800 Hz. If you do not see the waveform, ensure your power
source is adequate, the oscilloscope is powered- on, and the probe is connected securely
to the front- panel channel input BNC and to the probe calibration output.
5
Move the mouse around on the mouse surface and verify that the on- screen
mouse pointer coincides with the mouse movement.
Probe Compensation
Output
Содержание Infiniium 8000 Series
Страница 6: ...6...
Страница 31: ...Chapter 2 Preparing for Use Digital probe lead set 31 Data Sheets Demonstrations Catalogs...
Страница 66: ...Chapter 3 Testing Performance Performance Test Record 66...
Страница 140: ...Chapter 7 Replaceable Parts Exploded Views 140 Exploded Views Front Frame and Front Panel...
Страница 141: ...Chapter 7 Replaceable Parts Exploded Views 141 Fan and Acquisition Assembly...
Страница 142: ...Chapter 7 Replaceable Parts Exploded Views 142 Power Supply and PC Motherboard Without Option 017...
Страница 143: ...Chapter 7 Replaceable Parts Exploded Views 143 Power Supply and PC Motherboard With Option 017...
Страница 144: ...Chapter 7 Replaceable Parts Exploded Views 144 Sleeve and Accessory Pouch...
Страница 150: ...Chapter 8 Theory of Operation 150 Figure 8 1 Oscilloscope Block Diagram...
Страница 158: ...158...
Страница 160: ...160 Index...