Chapter 4: Calibrating and Adjusting
To run the self calibration
70
Figure 4-1
Calibration Dialog
3
Connect an adapter to each end of the cable.
4
Select Start, then follow the instructions on the screen.
You will be asked first to disconnect all channels and the Aux Output (located on the
rear panel of the oscilloscope), then to connect Aux Out to each channel and the
auxiliary trigger in turn. After you complete each cable change, select OK in the dialog
box to continue the calibration.
The oscilloscope displays a Passed/Failed message as each calibration routine is
completed.
5
Select Close.
Clear Cal Memory Protect to Run self calibration
You cannot run self calibration if this box is checked.
If calibration fails
Go to chapter 5 “Troubleshooting.”
Clear this check box
before starting
calibration
Select here to start
calibration
Содержание Infiniium 8000 Series
Страница 6: ...6...
Страница 31: ...Chapter 2 Preparing for Use Digital probe lead set 31 Data Sheets Demonstrations Catalogs...
Страница 66: ...Chapter 3 Testing Performance Performance Test Record 66...
Страница 140: ...Chapter 7 Replaceable Parts Exploded Views 140 Exploded Views Front Frame and Front Panel...
Страница 141: ...Chapter 7 Replaceable Parts Exploded Views 141 Fan and Acquisition Assembly...
Страница 142: ...Chapter 7 Replaceable Parts Exploded Views 142 Power Supply and PC Motherboard Without Option 017...
Страница 143: ...Chapter 7 Replaceable Parts Exploded Views 143 Power Supply and PC Motherboard With Option 017...
Страница 144: ...Chapter 7 Replaceable Parts Exploded Views 144 Sleeve and Accessory Pouch...
Страница 150: ...Chapter 8 Theory of Operation 150 Figure 8 1 Oscilloscope Block Diagram...
Страница 158: ...158...
Страница 160: ...160 Index...