Chapter 3: Testing Performance
To verify threshold accuracy
53
value of “Oscilloscope Threshold Voltage Setting” in Table 1 on page 55.
Close the Digital Setup dialog.
l
Set the horizontal time scale.
Select the horizontal scale setting (near the bottom of the display).
In the Enter Scale dialog, enter either 200 or 500 and then select p.
• For MSO/DSO8064A models, enter 500 and then select p.
• For MSO/DSO8104A models, enter 200 and then select p.
Select horizontal
scale setting