Chapter 6: Replacing Assemblies
To remove and replace the CD-ROM drive
102
To remove and replace the CD- ROM drive
Use this procedure to remove and replace the CD- ROM drive on units so equipped
(newer units will not have internal CD- ROM drives; an optional external DVD drive
will be available). When necessary, refer to other removal procedures.The graphics in
this chapter are representative of the oscilloscope at the time of this printing. Your
unit may look different.
1
Disconnect the power cable from the instrument and remove the top sleeve.
2
Remove the following cables:
• SATA Cable
• CD- ROM Power cable
3
Using a Torx T10 driver, remove the three CD- ROM support bracket screws
holding the CD- ROM drive in place.
When re- assembling, torque the three Torx T10 screws to 5 in- lb.
4
Using a T20 driver remove the back panel screw holding the CD- ROM drive in
place.
When re- assembling, torque the Torx T20 screw to 18 in- lb.
5
Move the CD- ROM drive back towards the front frame, then lift and pull to
remove.
Figure 6-6
Removing the CD-ROM assembly
Torx
T20
SATA cable
Power c
Torx
T10
Torx
T10
Torx
T10
Содержание Infiniium 8000 Series
Страница 6: ...6...
Страница 31: ...Chapter 2 Preparing for Use Digital probe lead set 31 Data Sheets Demonstrations Catalogs...
Страница 66: ...Chapter 3 Testing Performance Performance Test Record 66...
Страница 140: ...Chapter 7 Replaceable Parts Exploded Views 140 Exploded Views Front Frame and Front Panel...
Страница 141: ...Chapter 7 Replaceable Parts Exploded Views 141 Fan and Acquisition Assembly...
Страница 142: ...Chapter 7 Replaceable Parts Exploded Views 142 Power Supply and PC Motherboard Without Option 017...
Страница 143: ...Chapter 7 Replaceable Parts Exploded Views 143 Power Supply and PC Motherboard With Option 017...
Страница 144: ...Chapter 7 Replaceable Parts Exploded Views 144 Sleeve and Accessory Pouch...
Страница 150: ...Chapter 8 Theory of Operation 150 Figure 8 1 Oscilloscope Block Diagram...
Страница 158: ...158...
Страница 160: ...160 Index...