Chapter 3: Testing Performance
To test the DC calibrator
40
To test the DC calibrator
The Aux Out BNC on the back panel is used for self- calibration. Though calibrator
accuracy is not specified in the performance specifications, it must be within limits in
order to provide accurate self- calibration.
Test Limits:
- 2.4 v to +2.4 v, DC gain accuracy
±
0.2%
Equipment Required
Procedure
1
Connect the multimeter to the back panel Aux Out BNC.
Use the BNC cable and the BNC to banana plug adapter.
2
Press Default Setup to set the oscilloscope to default conditions.
3
Select Calibration from the Utilities menu.
4
Select DC from the Aux Output drop- down list box.
Figure 3-1
Selecting DC in the Calibration Dialog
Equipment
Critical Specifications
Recommended Model/Part
Digital Multimeter
0.1 mV resolution, better than 0.1% accuracy
Keysight 34401A
Cable
BNC
Keysight 10503A
Adapter
BNC (f) to banana (m)
Keysight 1251-2277
Set Aux Output to DC
Set the output voltage
Содержание Infiniium 8000 Series
Страница 6: ...6...
Страница 31: ...Chapter 2 Preparing for Use Digital probe lead set 31 Data Sheets Demonstrations Catalogs...
Страница 66: ...Chapter 3 Testing Performance Performance Test Record 66...
Страница 140: ...Chapter 7 Replaceable Parts Exploded Views 140 Exploded Views Front Frame and Front Panel...
Страница 141: ...Chapter 7 Replaceable Parts Exploded Views 141 Fan and Acquisition Assembly...
Страница 142: ...Chapter 7 Replaceable Parts Exploded Views 142 Power Supply and PC Motherboard Without Option 017...
Страница 143: ...Chapter 7 Replaceable Parts Exploded Views 143 Power Supply and PC Motherboard With Option 017...
Страница 144: ...Chapter 7 Replaceable Parts Exploded Views 144 Sleeve and Accessory Pouch...
Страница 150: ...Chapter 8 Theory of Operation 150 Figure 8 1 Oscilloscope Block Diagram...
Страница 158: ...158...
Страница 160: ...160 Index...