Key Reference
BW/SWP
87
VBW/RBW
Key access:
[BW/SWP] > {VBW/RBW}
Selects the ratio between the video and resolution
bandwidths.
If signal responses near the noise level are visually
masked by the noise, the ratio can be set to less than 1
to lower the noise. The knob and step keys change the
ratio in a 1, 3, 10 sequence.
Avg Type
Key access:
[BW/SWP] > {Avg Type}
Accesses the functions to automatically or manually
choose from one of the following averaging type:
Log Pwr
Key access:
[BW/SWP] > {Avg Type} > {Log Pwr}
Averages the data as appropriate for the logarithmic
scaled y-axis.
When average type is set to
Log Pwr
, “
LgAv
” appears on
the left side of the display.
Pwr
Key access:
[BW/SWP] > {Avg Type} > {Power}
Converts the scale from logarithmic to linear power
units, and then averages the power trace data.
With average type set to
Pwr
, “
PAvg
” appears on the left
side of the display.
NOTE
When preset is set to default, or VBW/RBW is set to Auto, the
ratio is set to 1.
Содержание N9340B
Страница 7: ...1 Overview 1 Overview ...
Страница 42: ...System and File File 36 ...
Страница 43: ...37 Making Measurements 4 Making Measurements ...
Страница 67: ...Making Measurements Making a Reflection Calibration Measurement 61 Figure 4 20 Short Circuit Normalized ...
Страница 150: ...Error Messages Error Message List 144 ...
Страница 154: ...Troubleshooting Contact Keysight Technologies 148 Contact Keysight Technologies ...
Страница 157: ...Menu Map AMPTD 151 ...
Страница 172: ...This information is subject to change without notice Keysight Technologies 2008 2015 Edition 3 December 2015 ...