Making Measurements
Making Distortion Measurements
53
After the performance of the source/analyzer
combination has been verified, the DUT (device under
test, for example, an amplifier) would be inserted
between the directional coupler output and the analyzer
input.
2 Set one source (signal generator) to 300 MHz and the
other source to 301 MHz. This will define the
frequency separation at 1 MHz. Set both sources
equal in amplitude, as measured by the analyzer. In
this example, they are both set to –5 dBm.
3 Set the analyzer center frequency and span:
•
Press [PRESET] (Factory preset)
•
Press [FREQ] > 300.5 > {MHz}
•
Press [SPAN] > 5 > {MHz}
4 Reduce the RBW until the distortion products are
visible:
•
Press [BW/SWP] > {RBW}, and reduce the RBW
using the knob, the arrow keys or the numeric key-
pad.
5 Move the signal to the reference level:
•
Press [MARKER] > {Peak Search}
•
Press [MARKER] > {Marker To} > {To Ref}
6 Reduce the RBW until the distortion products are
visible:
•
Press [BW/SWP] > {RBW}, and reduce the RBW
using the knob, the arrow keys or the numeric key-
pad.
7 Activate the second marker and place it on the peak
of the distortion product (beside the test signal) using
the Next Peak:
•
Press [MARKER] > {Delta}
NOTE
The coupler used should have a high isolation between the
two input ports to limit the sources intermodulation.
Содержание N9340B
Страница 7: ...1 Overview 1 Overview ...
Страница 42: ...System and File File 36 ...
Страница 43: ...37 Making Measurements 4 Making Measurements ...
Страница 67: ...Making Measurements Making a Reflection Calibration Measurement 61 Figure 4 20 Short Circuit Normalized ...
Страница 150: ...Error Messages Error Message List 144 ...
Страница 154: ...Troubleshooting Contact Keysight Technologies 148 Contact Keysight Technologies ...
Страница 157: ...Menu Map AMPTD 151 ...
Страница 172: ...This information is subject to change without notice Keysight Technologies 2008 2015 Edition 3 December 2015 ...