System and File
System Setting
31
rather than calibration course, as the calibration time is
unpredictable. When the calibration is finished, the LCD
displays a calibration, and the interrupted measurement
restarts.
Please refer to the operation procedures as below:
1 Use a BNC cable to connect a 10 MHz reference
signal to the EXT TRIG IN connector.
2 press [SYS] > {More} > {Calibration} > {Time Base} to
initiate a calibration.
Low Frequency Channel
Press [SYS] > {More} > {More} > {LowFreqChannel} to
enable the low frequency channel function. When this
softkey is enabled, the trace in the low frequency range
will has a better DANL.
This softkey is only available in spectrum analysis mode
with option XDM or option IBC.
NOTE
Time base calibration takes during a short time only when the
inner temperature is stable. When the inner temperature is
increasing, calibration takes a long-time course or will fail. If the
input reference signal is abnormal, the calibration cycle will take a
long and unpredictable time to exit, and the LCD displays an error
message.
Содержание N9340B
Страница 7: ...1 Overview 1 Overview ...
Страница 42: ...System and File File 36 ...
Страница 43: ...37 Making Measurements 4 Making Measurements ...
Страница 67: ...Making Measurements Making a Reflection Calibration Measurement 61 Figure 4 20 Short Circuit Normalized ...
Страница 150: ...Error Messages Error Message List 144 ...
Страница 154: ...Troubleshooting Contact Keysight Technologies 148 Contact Keysight Technologies ...
Страница 157: ...Menu Map AMPTD 151 ...
Страница 172: ...This information is subject to change without notice Keysight Technologies 2008 2015 Edition 3 December 2015 ...