Key Reference
MODE
118
Meas Setup
Key access:
[MODE] > {Power Meter} > {Meas Setup}
Accesses the Meas Disp submenu which is used to
configure the mesurement settings.
• Freq
Sets a frequency between 1 kHz and 90.0 GHz using
the numeric keypad, arrow key or knob. If the
frequency does not correspond directly to a frequency
in the sensor calibration table, the power sensor
calculates the calibration factor using linear
interpolation.
• Power Range
Accesses the
Power Range
submenu.
•
Auto Range
Enables and disable auto ranging. When auto
ranging is ON, the power sensor selects the best
measuring range for the measurement. When auto
ranging is OFF, the power sensor remains in the
current range and may be manually set by the user.
•
Range
Selects the Lower or Upper of power ranges when the
Auto Range is Off.
• Limits
Accesses the
Limits
submenu.
•
Limits
Enables or disables the limit function.
•
Upper Limit
Sets a value of the upper test limit for the current
measurement. The units used are dependent on the
current setting of power unit. When the measured
power is greater than the value specified, a failure will
be reported.
Содержание N9340B
Страница 7: ...1 Overview 1 Overview ...
Страница 42: ...System and File File 36 ...
Страница 43: ...37 Making Measurements 4 Making Measurements ...
Страница 67: ...Making Measurements Making a Reflection Calibration Measurement 61 Figure 4 20 Short Circuit Normalized ...
Страница 150: ...Error Messages Error Message List 144 ...
Страница 154: ...Troubleshooting Contact Keysight Technologies 148 Contact Keysight Technologies ...
Страница 157: ...Menu Map AMPTD 151 ...
Страница 172: ...This information is subject to change without notice Keysight Technologies 2008 2015 Edition 3 December 2015 ...