Making Measurements
Making an Average Power Measurement
67
Setting the Power Meter’s Upper and Lower
Limits
The internal Power Meter can be configured to detect
when a measurement has failed a user predefined upper
and lower limits.
1 Press [Preset] to perform a factory preset.
2 Press [MODE] > {Power Meter} > [ENTER] to turn on
the power meter.
3 Press {Meas Setup} > {Limits} to access the limits
menu.
4 Press {Limits} to activate the limits function. Each
time the softkey is pressed, the selected option
changes.
5 Press {Upper Limits} and enter the high limit value
using the numeric keypad. Then press {dBm} to
complete the setup.
6 Press {Lower Limits} and enter the low limit value
using the numeric keypad. Then press {dBm} to
complete the setup.