System and File
System Setting
30
Show System
Pressing [SYS] > {More} > {Show system} displays the
following hardware, software and battery information of
the analyzer:
Show Error
Pressing [SYS] > {More} > {Show Error} accesses a list of
the 30 most recent error messages. The most recent
error will appear at the bottom of the list. If the error list
is longer than 30 entries, the analyzer reports an error
message [–350, Query overflow]. When in remote
control, and the error list is greater than 30 entries, the
error display is halted and a message at the bottom of
the list informs the user that error exceed 30.
For more information, refer to
Perform a Time Base Calibration
The N9340B provides a manual calibration function to
calibrate the time base. The analyzer should warm up for
approximately 30 minutes before calibration.
When the calibration function is triggered, the current
measurement is interrupted and a gauge displays on the
LCD. The gauge simply indicates calibration action
Machine Model
Battery Info
MCU Firmware Version
Name
DSP Firmware Version
Serial NO.
FPGA Firmware Version
Capacity
RF Firmware Version
Temperature
RF Module S/N
Charge Cycles
KeyBoard Module S/N
Voltage
This Run Time
Current
Temperature
Charge Status
Source Voltage
Remain Time
Power Source
Host ID
Содержание N9340B
Страница 7: ...1 Overview 1 Overview ...
Страница 42: ...System and File File 36 ...
Страница 43: ...37 Making Measurements 4 Making Measurements ...
Страница 67: ...Making Measurements Making a Reflection Calibration Measurement 61 Figure 4 20 Short Circuit Normalized ...
Страница 150: ...Error Messages Error Message List 144 ...
Страница 154: ...Troubleshooting Contact Keysight Technologies 148 Contact Keysight Technologies ...
Страница 157: ...Menu Map AMPTD 151 ...
Страница 172: ...This information is subject to change without notice Keysight Technologies 2008 2015 Edition 3 December 2015 ...