Key Reference
Amplitude
84
HiSensitivity
Key access:
[AMPTD] > {HiSensitivity}
Press this softkey
On
to set 0 dB attenuation, -50 dBm
reference level and turn the pre-amplifier on. It’s very
helpful to observe the low-level signals with this
settings.
Ref Offset
Key access:
[AMPTD] > {Ref Offset}
Adds an offset value which ranges from -327.6 dB to
+327.6 dB, to the displayed reference level.
Reference-level offsets are used when gain or loss
occurs between a device under test and the analyzer
input. Thus, the signal level measured by the analyzer
may be referred to as the level at the input of an external
amplitude conversion device. When an amplitude offset
is entered, its value appears in the lower left corner of
the display.
Y Axis Units
Key access:
[AMPTD] > {More} > {Y Axis Units}
Sets the amplitude units.
NOTE
Reference-level offsets are only entered by using the numeric
keypad. Entering an offset does not affect the trace or the
attenuation value.
NOTE
Amplitude units are maintained for both logarithmic and linear
modes depending on the Scale Type setting.
Содержание N9340B
Страница 7: ...1 Overview 1 Overview ...
Страница 42: ...System and File File 36 ...
Страница 43: ...37 Making Measurements 4 Making Measurements ...
Страница 67: ...Making Measurements Making a Reflection Calibration Measurement 61 Figure 4 20 Short Circuit Normalized ...
Страница 150: ...Error Messages Error Message List 144 ...
Страница 154: ...Troubleshooting Contact Keysight Technologies 148 Contact Keysight Technologies ...
Страница 157: ...Menu Map AMPTD 151 ...
Страница 172: ...This information is subject to change without notice Keysight Technologies 2008 2015 Edition 3 December 2015 ...