Key Reference
MODE
119
•
Lower Limit
Sets the lower test limit value for the current mea-
surement. The units used are dependent on the cur-
rent setting of power unit. When the measured power
is less than the value specified, a failure will be
reported.
•
Limit Beep
Enables or disables the limit beep.
• Rel/Offset
Accesses the
Rel/Offset
submenu.
•
Rel
Enables the comparison of a measurement result to a
reference value and sets a reference value on the cur-
rent measured data.
•
Offset
Sets a value of the display offset on the current
measurement, and enable or disable the display
offset. The display offset is applied to the
measurement signal after any math calculation.
• Average
Accesses the
Average
submenu.
•
Average
Enables or disables the
Average
function.
•
Avg Count
Sets a value of the filter length and to set the average
count value to be Auto or Manual.
The power sensor automatically sets the number of
readings averaged together to satisfy most power
measurements averaging requirements. The number
of average readings depends on the resolution and
the power level settings the power sensor is currently
using.
•
Step Detect
Enables or disables step detection. In AUTO filter
mode, the average of the last four values entered into
the filter is compared to the entire filter average. If
Содержание N9340B
Страница 7: ...1 Overview 1 Overview ...
Страница 42: ...System and File File 36 ...
Страница 43: ...37 Making Measurements 4 Making Measurements ...
Страница 67: ...Making Measurements Making a Reflection Calibration Measurement 61 Figure 4 20 Short Circuit Normalized ...
Страница 150: ...Error Messages Error Message List 144 ...
Страница 154: ...Troubleshooting Contact Keysight Technologies 148 Contact Keysight Technologies ...
Страница 157: ...Menu Map AMPTD 151 ...
Страница 172: ...This information is subject to change without notice Keysight Technologies 2008 2015 Edition 3 December 2015 ...