System and File
System Setting
28
Upgrading Firmware
Follow the steps below to update the N9340B firmware:
1 Download the latest N9340B firmware from
2 Extract files to the root directory of a USB stick. You
will see a folder named “N9340DATA” with file
Bappupgrade.hy
.
3 Insert the USB stick into the top panel USB
connector.
4 Press [SYS] > {More} > {Upgrade} > {Firmware} to
activate the updating procedure. Press Enter to
upgrade the firmware. The analyzer will perform the
update automatically.
5 Unplug the USB stick and restart the N9340B when
message “All modules have been upgraded, please
restart” is displayed.
6 Press [SYS] > {More} > {Show System} to find the
updated MCU firmware version.
Ext Input
Toggles the channel for external input between Ref and
Trig. Ref refers to a 10 MHz reference signal; Trig refers
to a TTL signal.
NOTE
Please make sure your USB stick is in FAT16 or FAT32 format.
USB stick with self-startup partition or multi-partitions is not
support by N9340B.
CAU-
CAUTION
In updating process, there must be a constant power supply
to for at least 15 minutes. If power fails during the updating
process it can cause damage to the instrument.
Содержание N9340B
Страница 7: ...1 Overview 1 Overview ...
Страница 42: ...System and File File 36 ...
Страница 43: ...37 Making Measurements 4 Making Measurements ...
Страница 67: ...Making Measurements Making a Reflection Calibration Measurement 61 Figure 4 20 Short Circuit Normalized ...
Страница 150: ...Error Messages Error Message List 144 ...
Страница 154: ...Troubleshooting Contact Keysight Technologies 148 Contact Keysight Technologies ...
Страница 157: ...Menu Map AMPTD 151 ...
Страница 172: ...This information is subject to change without notice Keysight Technologies 2008 2015 Edition 3 December 2015 ...