Section 2: General operation
Models 707B and 708B Switching Matrix Reference Manual
2-78
707B-901-01 Rev. A / August 2010
The settings that are affected are:
•
Channels and patterns are removed from the scan list
•
Bypass: Returned to default setting of ON
•
Mode: Returned to default setting of Open All
•
Scan count: Returned to default setting of
1
•
Trigger to start scan: Set to Immediate
•
Trigger to continue channel action for each scan step: Setting is cleared
•
Arm (Scan Start Stimulus) is set to None
•
Channel Action Stimulus is set to Channel Ready Event
•
Channel Ready Event is set to None
•
Scan Complete Even is set to None
To reset the scan list from the web interface:
1. From the left navigation of the web interface Home page, select Scan Builder.
2. Click
Scan & Trigger Reset
.
Run the scan
You can run a scan in one of several ways:
•
Background:
Runs the scan in the background so that you can perform other tasks while the scan is
running. You can use the Query State to check scan status.
•
Step by Step:
Steps through the scan.
To run the scan as a background scan from the web interface:
Click
Execute Background
or
Step by Step
.
Stop the scan
To stop the scan from the web interface:
On the Build & Run tab, click
Abort
.
Monitor the state of the scan
To monitor the state of the scan, you can check the Query State box on the Build & Run tab. Query
State displays the current state of the scan, which can be:
•
Empty: No scan defined
•
Building: Scan list is being created
•
Running: Scan in process
•
Success: Scan completed successfully