Section 7: Command reference
Models 707B and 708B Switching Matrix Reference Manual
7-158
707B-901-01 Rev. A / August 2010
Example 1
scan.trigger.channel.stimulus =
scan.trigger.EVENT_SCAN_START
Sets the trigger stimulus of the channel
event detector to scan start event.
Example 2
scan.trigger.channel.stimulus = 0
Starts the channel action immediately
after the Scan Start Event.
Also see
(on page 7-156)
(on page 7-156)
scan.trigger.clear()
This function clears the trigger model.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function Yes
Usage
scan.trigger.clear()
Details
This function sets the arm and channel event detectors of the trigger model to the undetected state.
Example
scan.trigger.clear()
Clears the trigger model.
Also see
(on page 7-156)