Section 7: Command reference
Models 707B and 708B Switching Matrix Reference Manual
7-70
707B-901-01 Rev. A / August 2010
Key codes
Value Key
list
Value Key
list
0 display.KEY_NONE
83
display.KEY_RUN
66 display.KEY_DELETE
84
display.KEY_TRIG
67 display.KEY_EXIT
86
display.KEY_STEP
69 display.KEY_CLOSE
87
display.KEY_CHAN
70 display.KEY_SLOT
90
display.KEY_INSERT
72 display.KEY_DISPLAY
91
display.KEY_MENU
74 display.KEY_ENTER
93
display.KEY_OPEN
76 display.KEY_LOAD
94
display.KEY_PATT
77 display.KEY_SCAN
97
display.WHEEL_ENTER
79 display.KEY_OPENALL
107
display.WHEEL_LEFT
80 display.KEY_CONFIG
114
display.WHEEL_RIGHT
NOTE
When using this function, use built-in constants such as
display.KEY_STEP
(rather than the
numeric value of
86
). This will allow for better forward compatibility with firmware revisions.
Example
key = display.getlastkey()
print(key)
On the front panel, press the
MENU
key and
then send the code to the left. This retrieves the
key code for the last pressed key.
Output:
6.01
Also see
(on page 7-77)
display.gettext()
This function reads the text displayed on the instrument front panel.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function Yes