Section 7: TSP command reference
Model 2657A High Power System SourceMeter® Instrument Reference Manual
7-134
2657A-901-01 Rev. B/December 2012
Also see
(on page 7-126)
(on page 7-127)
lan.trigger[N].wait()
This function waits for an input trigger.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
triggered
= lan.trigger[
N
].wait(
timeout
)
triggered
Trigger detection indication
N
The trigger packet over LAN to wait for (1 to 8)
timeout
Maximum amount of time in seconds to wait for the trigger event
Details
If one or more trigger events have been detected since the last time
lan.trigger[
N
].wait()
or
lan.trigger[
N
].clear()
was called, this function returns immediately.
After waiting for a LAN trigger event with this function, the event detector is automatically reset and rearmed
regardless of the number of events detected.
Example
triggered = lan.trigger[5].wait(3)
Wait for a trigger with LAN packet 5 with a timeout of
3 seconds.
Also see
(on page 7-126)
(on page 7-132)
localnode.description
This attribute stores a user-defined description and mDNS service name of the instrument.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Not applicable
Nonvolatile memory
Instrument specific (see
Details
)
Usage
localnode.description =
description
description
= localnode.description
description
User-defined description and mDNS service name of the instrument; use a string of
63 characters or less