Details
Data for pulsed current measurements are stored in the reading buffer specified by the
buffer
input parameter.
This function configures a logarithmic pulsed voltage sweep with a current measurement at each point.
Measurements are made at the end of the
ton
time.
The magnitude of the first pulse will be
start
volts; the magnitude of the last pulse will be
stop
volts. The
magnitude of each pulse in between will be
LogStep
n
volts larger than the previous pulse, where:
LogStepSize = (log10(stop) - log10(start)) / (points -1)
LogStep
n
= (n - 1) * (LogStepSize), where n = [2, points]
SourceStepLevel
n
= antilog(LogStep
n
) * start
This function does not cause the specified
smu
to output a pulse train. It does check to see if all the pulse
dimensions can be achieved, and if they can, assigns the indicated
tag
or index to the pulse train.
Figure 114: ConfigPulseVMeasureISweepLog()
Example
ConfigPulseVMeasureISweepLog(smua, 0,
10, 1000, 10e-3, 20e-3, 100e-3, 10,
smua.nvbuffer1, 6)
Set up a pulsed logarithmic sweep that uses System
SourceMeter
®
instrument channel A. The pulsed
sweep starts at 10 V, ends at 1000 V, and returns to
a 0 V bias level between pulses. Each pulsed step is
on for 20 ms, and then at the bias level for 100 ms.
The current limit is 10 mA during the entire pulsed
sweep. The pulse train is comprised of 10 pulsed
steps, and the pulse train is assigned a
tag
index of
6.