Model 2657A High Power System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2657A-901-01 Rev. B/December 2012
7-63
display.getlastkey()
This function retrieves the key code for the last pressed key.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
keyCode
= display.getlastkey()
keyCode
A returned value that represents the last front-panel key pressed; see
Details
for
more information
Details
A history of the key code for the last pressed front-panel key is maintained by the instrument. When the
instrument is turned on, or when it is transitioning from local to remote operation, the key code is set to 0
(
display.KEY_NONE
).
Pressing the EXIT (LOCAL) key normally aborts a script. To use this function with the EXIT (LOCAL) key, you
must set
display.locallockout
to
display.LOCK
.
The table below lists the
keyCode
value for each front-panel action.
Key codes
Value Key
list
Value Key
list
0
display.KEY_NONE
82
display.KEY_ENTER
65
display.KEY_RANGEUP
85
display.KEY_RECALL
68
display.KEY_MENU
86
display.KEY_MEASA
69
display.KEY_MODEA
87
display.KEY_DIGITSA
70
display.KEY_RELA
92
display.KEY_TRIG
71
display.KEY_RUN
93
display.KEY_LIMITA
72
display.KEY_DISPLAY
94
display.KEY_SPEEDA
73
display.KEY_AUTO
95
display.KEY_LOAD
75
display.KEY_EXIT
97
display.WHEEL_ENTER
77
display.KEY_FILTERA
103
display.KEY_RIGHT
78
display.KEY_STORE
104
display.KEY_LEFT
79
display.KEY_SRCA
107
display.WHEEL_LEFT
80
display.KEY_CONFIG
114
display.WHEEL_RIGHT
81
display.KEY_RANGEDOWN
When using this function, use built-in constants, such as
display.KEY_RIGHT
, rather than the
numeric value, such as
103
. This allows for better forward compatibility with firmware revisions.
The OUTPUT ON/OFF control cannot be tracked by this function.
NOTE