Section 7: TSP command reference
Model 2657A High Power System SourceMeter® Instrument Reference Manual
7-206
2657A-901-01 Rev. B/December 2012
Usage
relEnable
= smu
X
.measure.rel.enable
Y
smu
X
.measure.rel.enable
Y
=
relEnable
relEnable
Relative measurement control. Set
relEnable
to one of the following values:
0
or
smu
X
.REL_OFF
: Disables relative measurements
1
or
smu
X
.REL_ON
: Enables relative measurements
X
Source-measure unit (SMU) channel (for example,
smua.measure.rel.enablev
applies to SMU channel A)
Y
SMU measurement function (
v
= voltage,
i
= current,
r
= resistance,
p
= power)
Details
This attribute enables or disables relative measurements. When relative measurements are enabled, all
subsequent measured readings are offset by the relative offset value specified by
smu
X
.measure.rel.level
Y
. Each returned measured relative reading is the result of the following
calculation:
Relative reading = Actual measured reading
−
Relative offset value
Example
smua.measure.rel.enablev = smua.REL_ON
Enables relative voltage measurements for SMU
channel A.
Also see
(on page 3-1)
(on page 7-211)