Model 2657A High Power System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2657A-901-01 Rev. B/December 2012
7-267
Bit
B7 B6 B5 B4 B3 B2 B1 B0
Binary
value
0/1 0/1 0/1 0/1 0/1 0/1 0/1 0/1
Decimal
128 64 32 16 8 4 2 1
Weights
(2
7
) (2
6
) (2
5
) (2
4
) (2
3
) (2
2
) (2
1
) (2
0
)
Bit
B15 B14 B13 B12 B11 B10 B9 B8
Binary
value
0/1 0/1 0/1 0/1 0/1 0/1 0/1 0/1
Decimal
32,768 16,384 8,192 4,096 2,048 1,024 512 256
Weights
(2
15
) (2
14
) (2
13
) (2
12
) (2
11
) (2
10
) (2
9
) (2
8
)
Example 1
operationRegister = status.operation.USER +
status.operation.PROG
status.operation.enable = operationRegister
Sets the USER and PROG bits of the
operation status enable register using
constants.
Example 2
-- decimal 20480 = binary 0101 0000 0000 0000
operationRegister = 20480
status.operation.enable = operationRegister
Sets the USER and PROG bits of the
operation status enable register using a
decimal value.
Also see
status.operation.calibrating.*
This attribute contains the operation status calibration summary register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
2 (All bits set)
Usage
operationRegister
= status.operation.calibrating.condition
operationRegister
= status.operation.calibrating.enable
operationRegister
= status.operation.calibrating.event
operationRegister
= status.operation.calibrating.ntr
operationRegister
= status.operation.calibrating.ptr
status.operation.calibrating.enable =
operationRegister
status.operation.calibrating.ntr =
operationRegister
status.operation.calibrating.ptr =
operationRegister
operationRegister
The status of the operation calibrating event register; a zero (0) indicates no bits set
(also send 0 to clear all bits); the only valid value other than 0 is 2