Model 2657A High Power System SourceMeter® Instrument Reference Manual
Section 2: General operation
2657A-901-01 Rev. B/December 2012
2-49
Item Description
Qty
Notes
(1)
Protective earth (safety ground) for Model
2657A
1
Keithley Instruments Model CA-568-120 is a protective
earth (safety ground) cable assembly supplied with the
instrument.
(2)
Interlock connection for Model 2657A
1
Pin 24 (INT) and Pin 22 (5 VDC) connected to test
fixture lid switch. Keithley Instruments Model 7709-308
is a 25-pin interlock male connector and can be used for
custom connections.
(3)
Interlocked metal safety enclosure
1
Select a safety enclosure with an interlock that has a
normally-open (NO) switch.
(4)
Protective earth (safety ground) for test
fixture
1
Redundant grounds may be required for specific test
setups (for example, setups that clamp LO).
(5)
Model HV-CA-571-3 high-voltage triaxial
female panel mount to unterminated cable
assembly
3
See schematic for connections. Make sure to follow
High-voltage triaxial cable termination
(6)
Model HV-CA-554 high-voltage triaxial
cable assembly (male to male)
3
(7)
Model 2657A High Power System
SourceMeter
®
instrument
1