
Section 7: TSP command reference
Model 2657A High Power System SourceMeter® Instrument Reference Manual
7-188
2657A-901-01 Rev. B/December 2012
Also see
(on page B-14)
(on page 7-184)
(on page 7-186)
smuX.contact.calibratelo()
This function calibrates the low/sense low contact check measurement.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
smu
X
.contact.calibratelo(
cp1Measured
,
cp1Reference
,
cp2Measured
,
cp2Reference
)
X
Source-measure unit (SMU) channel (for example,
smua.contact.calibratelo()
applies to SMU channel A)
cp1Measured
The value measured by this SMU for calibration point 1
cp1Reference
The reference measurement for calibration point 1 as measured externally
cp2Measured
The value measured by this SMU for calibration point 2
cp2Reference
The reference measurement for calibration point 2 as measured externally
Details
Contact check measurement calibration does not require range information.
Typically, calibration points one and two are near 0
Ω
and 50
Ω,
respectively.
All four measurements (
cp1Measured
,
cp1Reference
,
cp2Measured
, and
cp2Reference
) must be made
with the active calibration set. If not, corruption of the calibration constants may result.
The new calibration constants are activated immediately but are not written to nonvolatile memory. Use
smu
X
.cal.save()
to save the new constants to nonvolatile memory. The active calibration constants stay in
effect until the instrument is power cycled or a calibration set is loaded from nonvolatile memory with the
smu
X
.cal.restore()
function.
This function is disabled until a successful call to
smu
X
.cal.unlock()
is made.
Содержание 2657A
Страница 3: ......