Model 2657A High Power System SourceMeter® Instrument Reference Manual
Section 6: Instrument programming
2657A-901-01 Rev. B/December 2012
6-55
There are two methods you can use to run test scripts simultaneously:
•
Create multiple TSP-Link networks
•
Use a single TSP-Link network with groups
The following figure displays the first method, which consists of multiple TSP-Link networks. Each
TSP-Link network has a master node and a GPIB connection to the computer.
Figure 101: Model 2657A multiple TSP-Link networks
GPIB
TSP-Link In
TSP-Link Out
GPIB
TSP-Link In
TSP-Link Out
GPIB
TSP-Link In
TSP-Link Out
GPIB
TSP-Link In
TSP-Link Out
•
•
•
To PC
GPIB
TSP-Link In
TSP-Link Out
SMU 2
GPIB
TSP-Link In
TSP-Link Out
SMU 1
GPIB
TSP-Link In
TSP-Link Out
SMU 1
GPIB
TSP-Link In
TSP-Link Out
SMU 1
GPIB
TSP-Link In
TSP-Link Out
SMU 2
GPIB
TSP-Link In
TSP-Link Out
SMU
n
SMU
n
= last SMU in system.
1 Channel System
16+ Channel System
To PC
2 Channel System
To PC
Master
Subordinate
Master
Subordinate
Subordinate
Содержание 2657A
Страница 3: ......