Section 7: TSP command reference
Model 2657A High Power System SourceMeter® Instrument Reference Manual
7-152
2657A-901-01 Rev. B/December 2012
Figure 115: PulseIMeasureV
Example
PulseIMeasureV(smua, 0.001, 0.01,
20e-3, 40e-3, 10)
SMU A outputs 1 mA and dwells for 40 ms, outputs
10 mA and dwells for 20 ms. The voltage
measurements occur during each 20 ms dwell period.
After the measurement, the output returns to 1 mA
and dwells for 40 ms. This pulse-measure process
repeats nine more times.
Also see
PulseVMeasureI()
This KIPulse factory script function performs a specified number of pulse V, measure I cycles.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function No
Usage
PulseVMeasureI(
smu, bias, level, ton, toff, points
)
smu
System SourceMeter
®
instrument channel (set to
smua
)
bias
Bias level in volts
level
Pulse level in volts
ton
Pulse on time in seconds
toff
Pulse off time in seconds
points
Number of pulse-measure cycles
level
bias
ton
bias
toff
toff
Voltage measurement