Chapter 2:
Kaye ValProbe RT Hardware
Kaye ValProbe RT User Manual
12
Insulating Canister for High Temperatures
The Kaye ValProbe RT Insulating Canister (Part Number X2545) enables
you to operate bendable Temperature Loggers from up to 360°C by ensuring
that the Logger base does not exceed 140°C.
Figure 19: Insulating Canister (X2545)
IMPORTANT: Use the Insulating Canister in air applications only; do not
submerge the canister in liquid.
CAUTION!
Use the Insulating Canister in non-hazardous locations only.
Inserting a Logger into an Insulating Canister
For dual and single bendable Loggers only, unscrew the protective housing
cap from the Logger.
Remove the cap by pressing down on the cap and turning it counterclockwise
until the clips are free from the tabs.
Place the PEEK cylinder over the bendable Temperature Logger probe
and slide it to the base of the Logger.
Place the Logger and PEEK cylinder into the canister.
Insert the probe through the Insulating Canister cap center hole and slide
the cap to the cylinder base. Line up the clips with the tabs on the
cylinder base. Press down on the cap and turn clockwise so that the clips
are secured onto the tabs.
Figure 20: Inserting a Logger into an Insulating Canister
Содержание ValProbe RT
Страница 1: ...Validation Kaye ValProbe RT User Manual MXXX EN Rev 1 0 May 2020 ...
Страница 2: ... No content intended for this page ...
Страница 4: ... No content intended for this page ...
Страница 12: ......
Страница 147: ...Chapter 11 Qualification Reports Kaye ValProbe RT User Manual 134 No content intended for this page ...
Страница 163: ...Kaye ValProbe RT User Manual 150 ...
Страница 166: ... No content intended for this page ...
Страница 167: ... No content intended for this page ...