Chapter 10: Sensor Verification
Kaye ValProbe RT User Manual
95
For additional information on a Logger, select it and then press the
Logger Details
button. To
proceed with the Verification study, click
Sensor Information.
Figure 105: Loggers Listed
The Sensor Information screen
will allow you to do a last review of the loggers about to be
programmed. Ensure all Loggers selected are listed and then press
Program Loggers.
Figure 106: Loggers Listed
On the Program Loggers screen the Base Station receives the Verification Setup information
and sends it via RF to the previously selected Loggers. The programming of the Loggers will
be completed when all three status icons have turned green. Wait until completed and the
Initiate Verification blinks yellow and press
Initiate Verification
Figure 107: Program Loggers Screen
10.5. Running a Sensor Verification
Prior to starting the verification press the
Hardware
button and verify the IRTD and Kaye
bath drywell are displayed as connected.