Chapter 4: Creating User Accounts
Kaye ValProbe RT User Manual
43
4.5. Setting Preferences
Figure 53: Preferences Tab in the Admin Menu
Once the initial ValProbe RT software installs with default system settings, you can change the
settings inside the Preferences screen. The new settings become effective independently.
•
Company Name
: Enter a company name to display in report headers.
•
Temperature
: Use the drop-down to set the temperature units to Celsius or Fahrenheit. All
calculations are performed in the temperature units specified here.
Note:
After switching to a different temperature unit, setups must be recreated for the new
temperature unit.
•
Max Groups
: Use the drop-down to select a maximum number of groups for a setup.
•
Line Frequency
: Use the drop-down to select a line frequency, 60 Hertz or 50 Hertz depending
on your environment.
•
Pressure
: Use the drop-down to select the absolute pressure units for saturated steam
calculations. If you are selecting “Other”, enter the value of 1 Atmosphere in the absolute
pressure units you are using.
•
Alternative Machine ID
: If required, enter an alternate machine identification number, which
appears in the audit trail as additional identification.
•
Allow Users to change Lethality
: Use the drop-down to decide if the user should be able to
change lethality, Yes or No.
•
IRTD Stability Threshold Lower Limit
: Enter the lower limit for IRTD stability (from 250 to
400°C or 482 to 750°F). The IRTD stability is fixed to 0.012°C but for high temperatures which
threshold limit is set with this value, or temperatures below 0°C, the IRTD stability value
becomes editable in the setup.
•
Logo
: Press
Change
to browse to a location to add a logo to display in report headers. The
graphic can be any bitmap and the preferable size is 90 x 30 pixel. Other formats are downsized.
•
Report Footer
: Decide if signature fields should be applied on first/last/every page for all report
types. For detailed and summary reports, this is set in the report tool directly.
•
Data Directory
: Displays the path and folder holding the Data Directory.
Содержание ValProbe RT
Страница 1: ...Validation Kaye ValProbe RT User Manual MXXX EN Rev 1 0 May 2020 ...
Страница 2: ... No content intended for this page ...
Страница 4: ... No content intended for this page ...
Страница 12: ......
Страница 147: ...Chapter 11 Qualification Reports Kaye ValProbe RT User Manual 134 No content intended for this page ...
Страница 163: ...Kaye ValProbe RT User Manual 150 ...
Страница 166: ... No content intended for this page ...
Страница 167: ... No content intended for this page ...