Chapter
11: Qualification Reports
Kaye ValProbe RT User Manual
117
Equilibrium Duration
: The Equilibrium Duration calculates the time in seconds from when the
defined Reference Sensor in the Group reaches the Process Temperature until the last sensor
in the Group reaches the Process Temperature. The dropdown menu allows selection of the
reference sensor. The
Equilibrium Duration
field is only definable if the study has a one
second sampling rate. The criterion value is the Maximum time in seconds allowed. The
maximum definable criterion is 59 seconds. The criterion is “Passed” if the calculated
Equilibrium time is equal to or less than the defined criteria.
Exposure Duration/Holding Duration
: The Exposure / Holding Duration calculates the total
time of Exposure / Hold. This criterion is only displayed if the selected cycle is an exposure
cycle. A criterion for a minimum and a maximum time of Exposure/Hold can be defined
independently. The minimum time criterion “Passes” if the calculated Exposure/Hold time is
equal to or above the defined criteria. The maximum Exposure/Hold time criterion “Passes” if
the calculated time is equal to or below the defined criteria. If both Min and Max fields are
defined the criterion “Passes” if the calculated Exposure/Hold time is equal to or within the
defined range.
Содержание ValProbe RT
Страница 1: ...Validation Kaye ValProbe RT User Manual MXXX EN Rev 1 0 May 2020 ...
Страница 2: ... No content intended for this page ...
Страница 4: ... No content intended for this page ...
Страница 12: ......
Страница 147: ...Chapter 11 Qualification Reports Kaye ValProbe RT User Manual 134 No content intended for this page ...
Страница 163: ...Kaye ValProbe RT User Manual 150 ...
Страница 166: ... No content intended for this page ...
Страница 167: ... No content intended for this page ...