Chapter 10: Sensor Verification
Kaye ValProbe RT User Manual
92
Choose the
Temperature
verification type on the Select Verification Type window.
Figure 99: Select Verification Type
On the Temperature Verification Setup screen, the following options are available:
•
Choose between one to six temperature setpoints
•
Define for each setpoint the setpoint temperature, Pass/Fail criteria and results log
times for reports
•
Define sensor stability and minimum duration
•
IRTD temperature reference stability criteria and duration
•
Additional time at stability to ensure sensors, loggers, and references are stable
•
SOP protocol numbers or other information
Figure 100: Setup Temperature Verification
To enter additional Setpoints or delete setpoints click on the appropriate
icon. Once all
setpoints and criteria are entered, save the Verification Setup under a specific name. If
desired, setup printing is available by selecting the
Printer
icon. To delete a setup, click the
Trash Bin
icon.
Multiple Setups can be generated and saved with different configurations, all are listed on the
right of the screen by Setup name and sorted by creation date. To modify an existing setup,
double-click on it to open it, modify it and then click
Save
to save the changes or
Cancel
to
revert your changes.
Содержание ValProbe RT
Страница 1: ...Validation Kaye ValProbe RT User Manual MXXX EN Rev 1 0 May 2020 ...
Страница 2: ... No content intended for this page ...
Страница 4: ... No content intended for this page ...
Страница 12: ......
Страница 147: ...Chapter 11 Qualification Reports Kaye ValProbe RT User Manual 134 No content intended for this page ...
Страница 163: ...Kaye ValProbe RT User Manual 150 ...
Страница 166: ... No content intended for this page ...
Страница 167: ... No content intended for this page ...