Appendix A. Specifications
Kaye ValProbe RT User Manual
139
Bendable Loggers: till 400°C (on request – please contact factory).
A.4.5
Sensor Lengths
Rigid 1.5, 3, 6 and 9” inches. Bendable 12, 24 and 36” inches. Flexible 40” inches.
Please contact factory for special lengths.
A.5 Kaye ValProbe RT Std Temperature Logger
A.5.1
Measurement Accuracy
±0.1°C from -40 to 140°C
±0.25°C from 140 to 400°C for bendable models
A.5.2
Body Environmental ratings:
Temperature: -40 to 140°C.
Temperature: -40 to 360°C with canister model.
Humidity: 0 to 100%, condensing.
Pressure: 0 to 5 bar absolute.
A.5.3
Number of Input Channels
1 and 2 channels.
A.5.4
Sensor Types and Measurement Range
Rigid and flexible: -40 to 140°C.
Bendable: -40 to 400°C.
A.5.5
Sensor Lengths
Rigid 1.5, 3, 6 and 9” inches. Bendable 12, 24 and 36” inches. Flexible 12, 24 and 36”
inches. Please contact factory for special lengths.
A.6 Kaye ValProbe RT Pressure Temperature Logger
A.6.1
Measurement Accuracy
Temperature: ±0.1°C from 0 to 140°C.
Pressure: 0 to 5 bar absolute.
±25 mbar from 0 to 120°C.
±10 mbar from 120 to 135°C.
±25 mbar from 135 to 140°C.
Содержание ValProbe RT
Страница 1: ...Validation Kaye ValProbe RT User Manual MXXX EN Rev 1 0 May 2020 ...
Страница 2: ... No content intended for this page ...
Страница 4: ... No content intended for this page ...
Страница 12: ......
Страница 147: ...Chapter 11 Qualification Reports Kaye ValProbe RT User Manual 134 No content intended for this page ...
Страница 163: ...Kaye ValProbe RT User Manual 150 ...
Страница 166: ... No content intended for this page ...
Страница 167: ... No content intended for this page ...