manualshive.com logo in svg
background image

CALIBRATION PROCEDURE

NI 9263

4-Channel, ±10 V, 16-Bit Analog Voltage Output Module

This document contains the verification and adjustment procedures for the National Instruments 
9263. For more information on calibration solutions, visit 

ni.com/calibration

.

Contents

Software.................................................................................................................................... 1
Documentation.......................................................................................................................... 2
Test Equipment......................................................................................................................... 3
Test Conditions......................................................................................................................... 3
Initial Setup............................................................................................................................... 3
Verification............................................................................................................................... 4

Accuracy Verification....................................................................................................... 4

Adjustment................................................................................................................................ 6

Accuracy Adjustment ....................................................................................................... 6

EEPROM Update ..................................................................................................................... 6
Reverification ........................................................................................................................... 7
Accuracy Under Calibration Conditions .................................................................................. 7
Worldwide Support and Services ............................................................................................. 8

Software

Calibrating the NI 9263 requires the installation of NI-DAQmx 9.6 or later on the calibration 
system. You can download NI-DAQmx from 

ni.com/downloads

 NI-DAQmx supports 

LabVIEW, LabWindows

/CVI

, ANSI C, and .NET. When you install NI-DAQmx, you only 

need to install support for the application software that you intend to use.

Содержание NI 9263

Страница 1: ...Initial Setup 3 Verification 4 Accuracy Verification 4 Adjustment 6 Accuracy Adjustment 6 EEPROM Update 6 Reverification 7 Accuracy Under Calibration Conditions 7 Worldwide Support and Services 8 Sof...

Страница 2: ...ctions and Specifications NI 9263 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabVIEW progra...

Страница 3: ...NI 9263 board temperature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the NI 9263 measurement circuitry is...

Страница 4: ...Accuracy Verification Connections 3 Set the DMM to read DC voltage in the minimum range required to cover the NI 9263 output range of 10 V 4 Acquire a sample a Create and configure an AO voltage chann...

Страница 5: ...le was purchased before 2007 then use the Revision D limits Note Test limits in Table 4 are derived using the values in Table 7 Table 4 NI 9263 Revision D Test Limits Range V Test Point 1 Year Limits...

Страница 6: ...C value obtained from the array of recommended calibration voltages f Wait the appropriate amount of time for the DMM measurement to settle g Read the NI 9263 output voltage measurement from the DMM h...

Страница 7: ...e device to NI Accuracy Under Calibration Conditions The values in the following table are based on calibrated scaling coefficients which are stored in the onboard EEPROM The following accuracy table...

Страница 8: ...ur complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance from NI Ap...