Attention
Hold the new UVC lamp at one end (not the glass body) when removing it from the
package.
Carefully take the UVC lamp out of the package.
(see Chap. 5.2.3 for figures and procedure)
Carefully guide the UVC lamp (Fig. 18) into the quartz tube.
Connect the UVC lamp with the connection plug (Fig. 19) in accordance with the instructions, slide
the protective cover (Fig. 20) towards the radiation chamber as far as it will go and tighten the cable
connection.
Connect the protective conductor (Fig. 21) to the earth connection of the radiation chamber in
accordance with the instructions.
11.4 Cleaning the quartz tube
Warning
Risk of injury!
Before handling the quartz tube, ensure that the JUDO UV disinfection system is
hydraulically depressurised and completely drained!
Wear cut-resistant gloves when installing or removing the quartz tube (see Chap. 2.3.3
and Chap. 2.4)!
Warning
Make sure the quartz tube is not scratched. The UV disinfection system must not be
operated with a scratched quartz tube. A scratched quartz tube must be replaced
immediately.
The quartz tube must be cleaned regularly and whenever maintenance work is carried out or the “Low UVC
%” message is displayed.
Pull the plug from the mains supply.
Remove the protective conductor from the earth connection of the radiation chamber.
Unscrew the cable gland at the protective cover and slide the protective cover along the cable.
Carefully pull the connection plug from the UVC lamp.
Pull the UVC lamp out of the quartz tube and carefully set it down in a dry, grease-free area.
Pull the quartz tube carefully, slowly and in a straight line out of the radiation chamber.
Make sure the quartz tube is not dirty.