Summarized tests
During the recording of the identifiers, the display shows the relevant header. The
header changes as soon as the signal levels are measured.
Identifier scan
Status: scanning
Signal levels per id
Status: measuring
"Identifier scan" always lasts for the set scan time.
"Signal levels per identifier" lasts for the set maximum measuring time at the
longest (default setting 60 s).
When the differential and the absolute signal levels of all identifiers have been
measured as often as is required for the averaging (default setting 3 times), the
measurement is already ended before the set measuring time has elapsed.
Please note: it may take several minutes until the measurement is completed if
identifiers are transmitted at short intervals one after another (approx. 10 µs) or
the number of identifiers is large (maximum 2048).
In the case of correct signal levels, "TEST PASSED" is displayed. The tolerance
ranges for the differential and absolute signal level are given in section 10.1.
Signal levels per id
TEST PASSED
END INFO
End the operational tests with the virtual key "END". You can now see all test re-
sults under "Stored results".