SECTION 11: SPECIFICATIONS
Part No. 715-0080, Rev. B
- 93 -
Masimo Sensor Accuracy
4
During No Motion Conditions
5
Oxygen Saturation - Neonates ------------ 70 - 100%
3%
0 - 69%
unspecified
Oxygen Saturation - Pediatrics ------------ 70 - 100%
2%
0 - 69%
unspecified
Pulse Rate - Neonates / Pediatrics ------- 25 - 240 bpm
3 bpm
During Motion Conditions
6,7
Oxygen Saturation - Neonates / Pediatrics ------ 70 - 100%
3%
0 - 69%
unspecified
Pulse Rate - Neonates / Pediatrics ----------- 25 - 240 bpm
5 bpm
Low Perfusion (where 0.02% Pulse Amplitude and % Transmission > 5%)
8
Oxygen Saturation - Neonates / Pediatrics ------------------------
2%
Pulse Rate - Neonates / Pediatrics ----------------------------------
3 bpm
4
Masimo sensors have been validated for pulse rate accuracy for the range of 25 - 240 bpm in bench top testing against
a Biotek Index 2 simulator. This variation equals plus or minus one standard deviation, which encompasses 68% of
the population.
5
Masimo SET technology with LNCS sensors have been validated for no motion accuracy in human blood studies on
healthy adult male and female volunteers with light to dark skin pigmentation in induced hypoxia studies in the range of
70-100% SpO
2
against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard
deviation, which encompasses 68% of the population. The saturation accuracy of the neonatal sensors was validated
on adult male and female volunteers with light to dark skin pigmentation and 1% was added to account for the
properties of fetal hemoglobin.
6
Masimo SET technology with LNCS sensors has been validated for motion accuracy in human blood studies on
healthy adult male and female volunteers with light to dark skin pigmentation in induced hypoxia studies while
performing rubbing and tapping motions, at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between
1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO
2
against a laboratory
co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation, which encompasses 68%
of the population.
7
The Masimo SET Technology with LNCS Neo sensors has been validated for neonatal motion accuracy in human
blood studies on healthy adult volunteers in induced hypoxia studies while performing rubbing and tapping motions, at
2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in
induced hypoxia studies in the range of 70-100% SpO
2
against a laboratory CO-oximeter and ECG monitor. 1% has
been added to the saturation accuracy to account for the effects of fetal hemoglobin. This variation equals plus or
minus one standard deviation, which encompasses 68% of the population.
8
Masimo SET technology has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2
simulator and Masimo’s simulator with signal strengths of greater than 0.02% and a % transmission of greater than 5%
for saturations ranging from 70 to 100%. This variation equals plus or minus one standard deviation, which
encompasses 68% of the population.
Содержание AirBORNE A750i
Страница 1: ...A750i Transport Incubator Service Manual...
Страница 6: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 4 SCLAIM MER...
Страница 7: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 5 SCLAIM MER...
Страница 8: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 6 SCLAIM MER...
Страница 9: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 7 SCLAIM MER...
Страница 10: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 8 SCLAIM MER...
Страница 11: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 9 SCLAIM MER...
Страница 12: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 10 SCLAIM MER...
Страница 102: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B CONTROL SCHEM L BOARD S MATICS SCHEMAT DRAW TIC 001 51 WINGS 81 100...
Страница 103: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM PULSE MATICS OX DISPLA DRAW AY BOARD WINGS D SCHEMA ATIC 001 5 5193 101...
Страница 104: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 102...
Страница 105: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 103...
Страница 106: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 104...
Страница 107: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 105...
Страница 108: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 106...
Страница 110: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 108...
Страница 112: ...S Par ECTION rt No 715 008 SCHE N 14 S 80 Rev B MATIC PN SCHEM NEUMATIC MATICS C A750i CA DRAW ART 002 40 WINGS 010 110...