SECTION 4: PULSE OXIMETRY (OPTIONAL FEATURE)
Part No. 715-0080, Rev. B
- 45 -
SETTING HIGH AND LOW PULSE RATE ALARMS
NOTE:
The pulse oximeter display will convert back to Normal Operation Mode after 5
seconds if no buttons are pressed.
1.
Press the SET button (Figure 3 - 1, item 24) five times (or until the PULSE HI LED
illuminates) to set the high Pulse alarm.
2.
Use the up or down arrow buttons (Figure 3 - 1, item 25) to set the high alarm to the
desired setting.
a.
Must be at least 1 bpm higher than the Pulse low alarm setting.
b.
Can be set from 85 - 241 bpm.
3.
Press the SET button six times (or until the PULSE LO LED illuminates) to set the low
Pulse alarm.
4.
Use the up or down arrow buttons to set the low alarm to the desired setting.
a.
Must be at least 1 bpm lower than the Pulse high alarm setting.
b.
Can be set from 25 - 84 bpm.
SETTING THE ALGORITHM MODE (MASIMO ONLY)
NOTE:
The pulse oximeter display will convert back to Normal Operation Mode after 5
seconds if no buttons are pressed.
Different algorithm modes can be used depending on the importance of SpO
2
sensitivity vs.
probe-off detection. The system will power up in A1 Normal Mode as a default. When setting
the Algorithm Mode, the setting will be displayed on the %SPO2 screen.
1.
Press the SET button (Figure 3 - 1, item 24) 10 times (or until HI and LO %SPO2 LEDs
are both illuminated) to set the Algorithm Mode.
2.
Use the up or down arrow buttons (Figure 3 - 1, item 25) to choose one of the following
modes:
a.
A1 = Normal Mode
i.
Provides the best combination of sensitivity and probe-off detection
performance.
ii.
This mode is recommended for the majority of patients.
Содержание AirBORNE A750i
Страница 1: ...A750i Transport Incubator Service Manual...
Страница 6: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 4 SCLAIM MER...
Страница 7: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 5 SCLAIM MER...
Страница 8: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 6 SCLAIM MER...
Страница 9: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 7 SCLAIM MER...
Страница 10: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 8 SCLAIM MER...
Страница 11: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 9 SCLAIM MER...
Страница 12: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 10 SCLAIM MER...
Страница 102: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B CONTROL SCHEM L BOARD S MATICS SCHEMAT DRAW TIC 001 51 WINGS 81 100...
Страница 103: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM PULSE MATICS OX DISPLA DRAW AY BOARD WINGS D SCHEMA ATIC 001 5 5193 101...
Страница 104: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 102...
Страница 105: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 103...
Страница 106: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 104...
Страница 107: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 105...
Страница 108: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 106...
Страница 110: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 108...
Страница 112: ...S Par ECTION rt No 715 008 SCHE N 14 S 80 Rev B MATIC PN SCHEM NEUMATIC MATICS C A750i CA DRAW ART 002 40 WINGS 010 110...