SECTION 3: OPERATING INSTRUCTIONS
Part No. 715-0080, Rev. B
- 36 -
Power Interruption / Failure
If external AC or DC power is lost while the incubator is powered on, the incubator will automatically
switch over to battery power with no interruption in incubator performance or degradation of pulse
oximeter and/or oxygen monitor accuracy. In the event of total power loss, all user-programmed
settings (e.g. temperature setpoint, alarm limits) will default to the most recently programmed values.
POWERING UP THE INCUBATOR
To turn the incubator on, press the Power Button (Figure 3 - 1, item 32) on the lower right hand corner
of the Front Display Panel. When the incubator is turned on, all functional LEDs on the display panel
will illuminate and the audible alarm will sound. This Power-up Indicator Test will last approximately
two seconds after which the Main Display Screen will display the infant chamber air temperature set
point. The PulseOx Display screen will alternately display the HI/LO alarm settings for 16 seconds.
To interrupt the PulseOx display, press the PulseOx settings button. After the initial start-up, the
PulseOx display will be in the “Off” status. After 5 seconds, if no buttons have been pressed, the
display will change and show the actual measured infant chamber air temperature. The incubator will
immediately attempt to regulate the infant chamber air temperature to the last set point stored in
memory (refer to SECTION 3, SETTING THE INFANT CHAMBER AIR TEMPERATURE, for
instructions on setting the temperature).
One of three power source indicators will be illuminated, indicating which source is currently powering
the incubator.
AC power indicator (Figure 3 - 1, item 29)
External DC power indicator (Figure 3 - 1, item 30)
Internal battery power indicator (Figure 3 - 1, item 31)
POWERING OFF THE INCUBATOR
To turn the incubator off, press the Power Button (Figure 3 - 1, item 32) on the lower right hand corner
of the Front Display Panel. The Power Button provides the means of electrically isolating the incubator
circuits from supply mains simultaneously on all poles.
Содержание AirBORNE A750i
Страница 1: ...A750i Transport Incubator Service Manual...
Страница 6: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 4 SCLAIM MER...
Страница 7: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 5 SCLAIM MER...
Страница 8: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 6 SCLAIM MER...
Страница 9: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 7 SCLAIM MER...
Страница 10: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 8 SCLAIM MER...
Страница 11: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 9 SCLAIM MER...
Страница 12: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 10 SCLAIM MER...
Страница 102: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B CONTROL SCHEM L BOARD S MATICS SCHEMAT DRAW TIC 001 51 WINGS 81 100...
Страница 103: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM PULSE MATICS OX DISPLA DRAW AY BOARD WINGS D SCHEMA ATIC 001 5 5193 101...
Страница 104: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 102...
Страница 105: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 103...
Страница 106: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 104...
Страница 107: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 105...
Страница 108: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 106...
Страница 110: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 108...
Страница 112: ...S Par ECTION rt No 715 008 SCHE N 14 S 80 Rev B MATIC PN SCHEM NEUMATIC MATICS C A750i CA DRAW ART 002 40 WINGS 010 110...