SECTION 7: TROUBLESHOOTING
Part No. 715-0080, Rev. B
- 65 -
MAIN DISPLAY ERROR CODES
Table 7 - 1 shows the failure codes that may be seen on the Main Display Screen. If further
assistance is needed, contact International Biomedical.
Table 7 - 1 Main Display Screen Failure Codes
Error
Code
Failure
Corrective Action
E01
Set Point Output Voltage Error
The display board needs to be
replaced. Contact International
Biomedical for assistance.
E02
Temperature Out of Range: baby temp.
probe is disconnected, damaged, or out
of sensor’s useable range
Reconnect temperature probe to
incubator. If error persists,
replace temperature probe.
E03
Battery Voltage Out of Range: battery is
drained or overcharged
Charge battery by plugging into
AC power for a minimum of 8
hours. If problem persists,
battery must be replaced.
E04
Display Range Limit Exceeded: value to
be shown is outside of the display’s
limits
Verify temperature using
secondary method. If
temperature seems reasonable,
the temperature sensor is faulty
and must be replaced. Contact
International Biomedical for
assistance
E07
EEPROM Write Failure: temperature set
point not saved
Turn the incubator off and then
on again. Repeat temperature
set procedure. If error still
occurs, the display board must
be serviced. Contact
International Biomedical for
assistance.
E08
EEPROM Value Out of Range:
temperature set point out of range
E09 EEPROM
Corrupt
PULSE OXIMETER AND OXYGEN MONITOR FAILURE CODES
(OPTIONAL FEATURE)
Table 7 - 2 shows the failure codes that may be seen when using the Masimo Pulse Oximeter. Failure
codes will be displayed in the %SPO2, PULSE, or OXYGEN display screens, as designated in the
table. In some cases, a power down may reset the error; however, the pulse oximeter and/or oxygen
monitor should be used with extreme caution. If further assistance is needed, contact International
Biomedical.
Содержание AirBORNE A750i
Страница 1: ...A750i Transport Incubator Service Manual...
Страница 6: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 4 SCLAIM MER...
Страница 7: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 5 SCLAIM MER...
Страница 8: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 6 SCLAIM MER...
Страница 9: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 7 SCLAIM MER...
Страница 10: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 8 SCLAIM MER...
Страница 11: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 9 SCLAIM MER...
Страница 12: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 10 SCLAIM MER...
Страница 102: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B CONTROL SCHEM L BOARD S MATICS SCHEMAT DRAW TIC 001 51 WINGS 81 100...
Страница 103: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM PULSE MATICS OX DISPLA DRAW AY BOARD WINGS D SCHEMA ATIC 001 5 5193 101...
Страница 104: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 102...
Страница 105: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 103...
Страница 106: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 104...
Страница 107: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 105...
Страница 108: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 106...
Страница 110: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 108...
Страница 112: ...S Par ECTION rt No 715 008 SCHE N 14 S 80 Rev B MATIC PN SCHEM NEUMATIC MATICS C A750i CA DRAW ART 002 40 WINGS 010 110...