SECTION 6: PREVENTATIVE MAINTENANCE
Part No. 715-0080, Rev. B
- 61 -
3.
Inspect tanks for abrasion, cuts, or other damage.
4.
The tanks should be replaced according to the manufacturer’s recommendation.
INFANT CHAMBER INSPECTION & CLEANING
1.
Carefully inspect the infant chamber for cracks and crazing of the Plexiglas.
2.
Check all screws and knobs for tightness.
CAUTION:
Do not overtighten the infant chamber screws!
3.
Wash chamber with mild soap or detergent and water solution to clean the Plexiglas.
Household ammonia in water in the concentrations recommended for hospital cleaning is
also excellent. These solutions may be applied with a soft absorbent cloth, followed by
rinsing with clean water. Wipe away residue with soft chamois or cellulose sponge
material.
CAUTION:
Any type of alcohol or solvents should never be used for cleaning the
infant chamber, as these solvents tend to soften the surface and often
cause crazing. Cleansing materials containing abrasives or strong
solvents as found in some window cleaning preparations should never
be used to clean the inner and outer infant chamber.
BATTERY TEST
The incubator is equipped with a 12 volt 26 ampere-hour battery. Under normal operating
conditions in an ambient temperature of 20
C, a new fully charged battery should operate the
incubator for over three hours. If the operating time is less than 1 ¼ hours, the battery needs to
be replaced.
The battery is specified for only a finite number of complete discharges. The test of the battery
will completely discharge it. The test can be performed as follows:
1.
Ensure the battery is fully charged by connecting the incubator to AC power for 12 hours
with the power switch off.
2.
Place the incubator in a room with a temperature of approximately 20 - 23
C. Turn on
the incubator power switch and adjust the setpoint on the front panel to 37
C. Let the
incubator stabilize to its operational temperature with AC power.
3.
After the incubator has reached operating temperature, allow an additional twenty
minutes for stabilization.
Содержание AirBORNE A750i
Страница 1: ...A750i Transport Incubator Service Manual...
Страница 6: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 4 SCLAIM MER...
Страница 7: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 5 SCLAIM MER...
Страница 8: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 6 SCLAIM MER...
Страница 9: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 7 SCLAIM MER...
Страница 10: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 8 SCLAIM MER...
Страница 11: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 9 SCLAIM MER...
Страница 12: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 10 SCLAIM MER...
Страница 102: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B CONTROL SCHEM L BOARD S MATICS SCHEMAT DRAW TIC 001 51 WINGS 81 100...
Страница 103: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM PULSE MATICS OX DISPLA DRAW AY BOARD WINGS D SCHEMA ATIC 001 5 5193 101...
Страница 104: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 102...
Страница 105: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 103...
Страница 106: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 104...
Страница 107: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 105...
Страница 108: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 106...
Страница 110: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 108...
Страница 112: ...S Par ECTION rt No 715 008 SCHE N 14 S 80 Rev B MATIC PN SCHEM NEUMATIC MATICS C A750i CA DRAW ART 002 40 WINGS 010 110...