SECTION 4: PULSE OXIMETRY (OPTIONAL FEATURE)
Part No. 715-0080, Rev. B
- 47 -
SETTING THE PERFUSION INDEX (MASIMO ONLY)
The Perfusion Index value represents a ratio between the pulse signal and noise. It helps
clinicians determine if the pulse oximetry sensor is placed on an optimal monitoring site. The
Perfusion Index value has a range of 0.0 - 20.0 (i.e. the higher the value, the better the
perfusion).
1.
Press and hold the down arrow button (Figure 3 - 1, item 25).
2.
The perfusion index will be displayed on the %SPO2 screen. A display of “---“ indicates
no value is available.
3.
Release the down arrow button to switch back to Normal Operation Mode.
NOTE:
Calibration and resetting the pulse oximeter algorithm or averaging mode
may cause a slight delay in readings.
NOTE:
Setting alarm set points, calibration, and resetting pulse oximeter algorithm
or averaging methods will disable a portion of the pulse oximetry or oxygen
monitoring displays (Masimo only).
OPERATIONAL ALARMS AND WARNINGS
When the measured oxygen saturation or pulse rate falls out of the user-selected range, either “HI” or
“LO” will flash on the appropriate screen approximately every 2 to 3 seconds. An audible alarm will
also be heard every 3 seconds. Pressing the
MUTE
button will silence the audible alarm for
approximately 1 minute.
Update Period Exceeded
The PulseOx system updates the data being displayed on the display screens every second. After a
period of 5 seconds, if the system has not received an update, the screen will flash “PS”. If an update
still is unavailable after 15 seconds, “PS” will continue to flash and an audible alarm will sound until the
condition is corrected.
Содержание AirBORNE A750i
Страница 1: ...A750i Transport Incubator Service Manual...
Страница 6: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 4 SCLAIM MER...
Страница 7: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 5 SCLAIM MER...
Страница 8: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 6 SCLAIM MER...
Страница 9: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 7 SCLAIM MER...
Страница 10: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 8 SCLAIM MER...
Страница 11: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 9 SCLAIM MER...
Страница 12: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 10 SCLAIM MER...
Страница 102: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B CONTROL SCHEM L BOARD S MATICS SCHEMAT DRAW TIC 001 51 WINGS 81 100...
Страница 103: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM PULSE MATICS OX DISPLA DRAW AY BOARD WINGS D SCHEMA ATIC 001 5 5193 101...
Страница 104: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 102...
Страница 105: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 103...
Страница 106: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 104...
Страница 107: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 105...
Страница 108: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 106...
Страница 110: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 108...
Страница 112: ...S Par ECTION rt No 715 008 SCHE N 14 S 80 Rev B MATIC PN SCHEM NEUMATIC MATICS C A750i CA DRAW ART 002 40 WINGS 010 110...