SECTION 4: PULSE OXIMETRY (OPTIONAL FEATURE)
Part No. 715-0080, Rev. B
- 46 -
b.
A2 = Maximum Mode
i.
Interprets and displays data for very weak signals.
ii.
This mode is recommended for the sickest patients or during procedures
when clinician and patient contact is continuous.
c.
A3 = APOD Mode
i.
Least sensitive in picking up a reading on patients with low perfusion but
provides the best detection for probe-off conditions.
ii.
This mode is recommended for use on patients that are at particular risk of
the sensor becoming detached (e.g. pediatric, combative, etc.)
SETTING THE AVERAGING MODE (MASIMO ONLY)
NOTE:
The pulse oximeter display will convert back to Normal Operation Mode after 5
seconds if no buttons are pressed.
The Averaging Mode is used to determine how many values will be used to calculate the SpO
2
and pulse rate readings. A lower averaging rate uses fewer values to calculate the readings,
resulting in more unstable readings, but will more quickly alert the user to potentially alarming
patient conditions. The system will power up with C2 Middle Averaging Rate as a default.
When setting the Averaging Mode, the setting will be displayed on the %SPO2 screen.
1.
Press the SET button (Figure 3 - 1, item 24) 11 times (or until the HI and LO PULSE
LEDs are both illuminated) to set the Averaging Mode.
2.
Use the up or down arrow buttons (Figure 3 - 1, item 25) to choose one of the following
averaging rates:
a.
C1 = Low Averaging Rate
i. SpO
2
averaging rate is set to 2 - 4 seconds.
b.
C2 = Middle Averaging Rate
i. SpO
2
averaging rate is set to 8 seconds.
c.
C3 = High Averaging Rate
i. SpO
2
averaging rate is set to 16 seconds.
Содержание AirBORNE A750i
Страница 1: ...A750i Transport Incubator Service Manual...
Страница 6: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 4 SCLAIM MER...
Страница 7: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 5 SCLAIM MER...
Страница 8: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 6 SCLAIM MER...
Страница 9: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 7 SCLAIM MER...
Страница 10: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 8 SCLAIM MER...
Страница 11: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 9 SCLAIM MER...
Страница 12: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 10 SCLAIM MER...
Страница 102: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B CONTROL SCHEM L BOARD S MATICS SCHEMAT DRAW TIC 001 51 WINGS 81 100...
Страница 103: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM PULSE MATICS OX DISPLA DRAW AY BOARD WINGS D SCHEMA ATIC 001 5 5193 101...
Страница 104: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 102...
Страница 105: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 103...
Страница 106: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 104...
Страница 107: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 105...
Страница 108: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 106...
Страница 110: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 108...
Страница 112: ...S Par ECTION rt No 715 008 SCHE N 14 S 80 Rev B MATIC PN SCHEM NEUMATIC MATICS C A750i CA DRAW ART 002 40 WINGS 010 110...