Part No. 715-0080, Rev. B
- 56 -
SECTION 6: PREVENTATIVE MAINTENANCE
To ensure proper operation, standby readiness, and malfunction reporting, International Biomedical
recommends following a preventative maintenance program. The daily preventative maintenance
procedures can be performed by a knowledgeable incubator transport person. Monthly maintenance
procedures should be performed by an appropriately trained biomedical maintenance person. No
other preventative maintenance is required.
NOTE:
Perform preventative maintenance procedures on all Accessory Equipment as
recommended in the manual for each piece of equipment.
DAILY MAINTENANCE
The following maintenance procedures should be performed on a daily basis by a knowledgeable
incubator transport person.
OPERATIONAL CHECK
1.
Check the AC power cord and ensure that there are no cuts or severe bends in the cord,
that all three prongs on the plug are in good condition, and that the cord is securely
fastened in place. Replace the cord as needed.
2.
Ensure the incubator is plugged in to an appropriate AC power source. Verify the
BAT
CHG
indicator (Figure 3 - 1, item 28) on the front panel is illuminated and that all other
indicators are off when the incubator power is off.
3.
With the incubator connected to AC power, press the power button (Figure 3 - 1, item 32)
to turn the incubator on. The incubator will perform a test of all indicators, LEDs, and the
audible alarm. Ensure all the indicators and LEDs illuminate, and that the audible alarm
can be heard. Verify the
AC OP
indicator (Figure 3 - 1, item 29) remains illuminated
after completion of the test.
4.
Place a hand inside the right side of the infant chamber and verify air flow. The fan
should be able to be heard at the left side of the incubator.
5.
A few minutes after turning the incubator on, the air temperature inside the infant
chamber should begin to increase and will be displayed on the Main Display Screen.
Verify.
6.
Turn on the external light and ensure it operates under AC power. Turn off the light.
Содержание AirBORNE A750i
Страница 1: ...A750i Transport Incubator Service Manual...
Страница 6: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 4 SCLAIM MER...
Страница 7: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 5 SCLAIM MER...
Страница 8: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 6 SCLAIM MER...
Страница 9: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 7 SCLAIM MER...
Страница 10: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 8 SCLAIM MER...
Страница 11: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 9 SCLAIM MER...
Страница 12: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 10 SCLAIM MER...
Страница 102: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B CONTROL SCHEM L BOARD S MATICS SCHEMAT DRAW TIC 001 51 WINGS 81 100...
Страница 103: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM PULSE MATICS OX DISPLA DRAW AY BOARD WINGS D SCHEMA ATIC 001 5 5193 101...
Страница 104: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 102...
Страница 105: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 103...
Страница 106: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 104...
Страница 107: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 105...
Страница 108: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 106...
Страница 110: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 108...
Страница 112: ...S Par ECTION rt No 715 008 SCHE N 14 S 80 Rev B MATIC PN SCHEM NEUMATIC MATICS C A750i CA DRAW ART 002 40 WINGS 010 110...