SECTION 6: PREVENTATIVE MAINTENANCE
Part No. 715-0080, Rev. B
- 58 -
INFANT CHAMBER CHECK
CAUTION:
Do not move the incubator by pushing on the infant chamber. The infant
chamber is not designed to sustain the forces to push the incubator. Stress
fractures in the infant chamber can occur.
1.
Ensure that the infant chamber is free from cracks.
2.
Ensure that the gasket between the chambers and airflow tray is in place. Check that
the grommets in the tubing through-holes on the infant chamber are also in place.
3.
Ensure that the infant chamber is secured to the incubator with the two latches on both
ends of the infant chamber.
CAUTION:
Do not overtighten the infant chamber screws. Do not strip the
aluminum into which these screws are threaded.
4.
Ensure that the infant chamber is clean and ready for transport, as determined by the
transport team leader.
WARNING:
This incubator was calibrated with the infant chamber originally
supplied. If this chamber is exchanged for an infant chamber of
a different configuration or size, the temperature calibration will
be affected. Consult International Biomedical before returning
the incubator to service.
CART AND ACCESSORIES
1.
Inspect the casters for wear or damage. Inspect for worn or damaged rubber and
tightness of retaining nuts. Replace damaged casters immediately.
2.
Ensure the incubator is securely fastened to the cart.
3.
Test operation of other accessories as indicated by manufacturer’s recommendations.
CAUTION:
When the incubator is attached to a rigid cart with an internal power
strip, the power cord must be contained within cart or secured on the
end of cart with cord holders when not in use.
SENSORS AND CABLES (OPTIONAL EQUIPMENT)
1.
Inspect the oxygen sensor and the pulse oximeter sensor for damage. Replace, if
necessary.
2.
Inspect the oxygen sensor cable and pulse oximeter cable for wear or damage.
Replace, if necessary.
Содержание AirBORNE A750i
Страница 1: ...A750i Transport Incubator Service Manual...
Страница 6: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 4 SCLAIM MER...
Страница 7: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 5 SCLAIM MER...
Страница 8: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 6 SCLAIM MER...
Страница 9: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 7 SCLAIM MER...
Страница 10: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 8 SCLAIM MER...
Страница 11: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 9 SCLAIM MER...
Страница 12: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 10 SCLAIM MER...
Страница 102: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B CONTROL SCHEM L BOARD S MATICS SCHEMAT DRAW TIC 001 51 WINGS 81 100...
Страница 103: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM PULSE MATICS OX DISPLA DRAW AY BOARD WINGS D SCHEMA ATIC 001 5 5193 101...
Страница 104: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 102...
Страница 105: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 103...
Страница 106: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 104...
Страница 107: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 105...
Страница 108: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 106...
Страница 110: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 108...
Страница 112: ...S Par ECTION rt No 715 008 SCHE N 14 S 80 Rev B MATIC PN SCHEM NEUMATIC MATICS C A750i CA DRAW ART 002 40 WINGS 010 110...